Dealer: This manual MUST be given to the end user.User: Before using this product, read this manual and save for future reference.Operator’s Manual Fo
SECTION 1—GENERAL GUIDELINESPlatinum™ Series 10 Part No 1118353Otherdevicesmayexperienceinterferencefromeventhelowlevelsofelectromagnetice
SECTION 2—FEATURESPart No 1118353 11 Platinum™ SeriesSECTION 2—FEATURESOxygen OutletFlowmeterOxygen Purity Indicator Lights (SensO2 model only)/ Fault
SECTION 3—HANDLINGPlatinum™ Series 12 Part No 1118353SECTION 3—HANDLINGTheconcentratorshouldALWAYSbekeptintheuprightpositiontopreventcabin
SECTION 4—TYPICAL PRODUCT PARAMETERSPart No 1118353 13 Platinum™ SeriesSECTION 4—TYPICAL PRODUCT PARAMETERS*NOTE:Statedconcentrationlevelsachieved
SECTION 4—TYPICAL PRODUCT PARAMETERSPlatinum™ Series 14 Part No 1118353Flow Range: Platinum 5 and XL - 0.5 to 5 L/min For flowrates less than 1 L/min,
SECTION 4—TYPICAL PRODUCT PARAMETERSPart No 1118353 15 Platinum™ SeriesOperating Ambient Temperature:50°F - 95°F (10°C - 35°C) at 20-60% relative humi
SECTION 5—OPERATING INSTRUCTIONSPlatinum™ Series 16 Part No 1118353SECTION 5—OPERATING INSTRUCTIONSIntroductionYouroxygenconcentratorisintendedfo
SECTION 5—OPERATING INSTRUCTIONSPart No 1118353 17 Platinum™ SeriesSet-UpPlug in Power CordPluginpowercordtoanelectricaloutlet. POLARIZED PLUG
SECTION 5—OPERATING INSTRUCTIONSPlatinum™ Series 18 Part No 1118353FIGURE 5.2 Humidifier Compartment3. Removethefilteraccessdoorlocatedonthe
SECTION 5—OPERATING INSTRUCTIONSPart No 1118353 19 Platinum™ Series7. Attachoxygentubingfromthehumidifierbottletooxygenoutletconnectoronth
WARNINGPlatinum™ Series 2 Part No 1118353 WARNINGDONOTusethisproductoranyavailableoptionalequipmentwithoutfirstcompletelyreadingandund
SECTION 5—OPERATING INSTRUCTIONSPlatinum™ Series 20 Part No 1118353FlowrateNOTE:Forthisprocedure,refertoFIGURE5.6onpage 20.1. Turntheflowra
SECTION 5—OPERATING INSTRUCTIONSPart No 1118353 21 Platinum™ SeriesFront Panel Indicators - Units with SensO2NOTE:Forthisprocedure,refertoFIGURE
SECTION 5—OPERATING INSTRUCTIONSPlatinum™ Series 22 Part No 1118353Front Panel Indicators - Units without SensO2NOTE:Forthisprocedure,refertoFIG
SECTION 6—MAINTENANCEPart No 1118353 23 Platinum™ SeriesSECTION 6—MAINTENANCE WARNINGTheInvacareconcentratorsarespecificallydesignedtominimize
SECTION 6—MAINTENANCEPlatinum™ Series 24 Part No 11183532. Cleanthecabinetfilterswithavacuumcleanerorwashinwarmsoapywaterandrinsethoro
SECTION 6—MAINTENANCEPart No 1118353 25 Platinum™ SeriesFIGURE 6.2 Preventive Maintenance RecordON EACH INSPECTIONMODEL NO. IRC SERIAL NO.PREVEN
SECTION 7—TROUBLESHOOTING GUIDEPlatinum™ Series 26 Part No 1118353SECTION 7—TROUBLESHOOTING GUIDESYMPTOM: PROBABLE CAUSE: SOLUTION:Alarm:SHORT BEEPS,
SECTION 7—TROUBLESHOOTING GUIDEPart No 1118353 27 Platinum™ SeriesYELLOW or RED LIGHT Illuminated.Alarm:CONTINUOUSOn red light only.* Only applies to
NOTESPlatinum™ Series 28 Part No 1118353NOTES
NOTESPart No 1118353 29 Platinum™ SeriesNOTES
Part No 1118353 3 Platinum™ SeriesTABLE OF CONTENTSTABLE OF CONTENTSACCESSORIES ... 2REGISTER YOUR PROD
LIMITED WARRANTYPlatinum™ Series 30 Part No 1118353LIMITED WARRANTYNOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO
LIMITED WARRANTYPart No 1118353 31 Platinum™ SeriesFor warranty service, please contact Invacare Corporation's Service Department at the toll fre
Invacare Corporationwww.invacare.comUSAOne Invacare WayElyria, Ohio USA44036-2125800-333-6900Technical Services800-832-4707Part No 1118353Trademarks a
TABLE OF CONTENTSPlatinum™ Series 4 Part No 1118353TABLE OF CONTENTSSECTION 6—MAINTENANCE ... 23Routine Maintenance...
SPECIAL NOTESPart No 1118353 5 Platinum™ SeriesSPECIAL NOTESSignalwordsare usedinthismanualandapplytohazardsorunsafepracticeswhichcould
SPECIAL NOTESPlatinum™ Series 6 Part No 1118353CAUTION“Caution:Federallawrestrictsthisdevicetosaleorrentalbyoronorderofaphysician,or
LABEL LOCATIONPart No 1118353 7 Platinum™SeriesLABEL LOCATIONHomeFill II Compatib leI/OOperating Instructions(See Operator's Manual)Alarms(See
SECTION 1—GENERAL GUIDELINESPlatinum™ Series 8 Part No 1118353SECTION 1—GENERAL GUIDELINESInordertoensurethesafeinstallation,assemblyandopera
SECTION 1—GENERAL GUIDELINESPart No 1118353 9 Platinum™ SeriesOperating InformationForoptimumperformance,Invacarerecommendsthateachconcentrator