User ManualThis manual MUST be given to the user of the product.BEFORE using this product, read this manual and save for future reference.ENInvacare®M
3 SAFETYInvacare®MA80 Series Mattresses 10 Part No 11481383 Safety3.1 General GuidelinesContraindicationsWARNINGDO NOT use this product or any avai
3 SAFETYPart No 1148138 11 Invacare®MA80 Series MattressesInstallationWARNINGThe MA80 series and MA85 series systems (including bariatric and raised
3 SAFETYInvacare®MA80 Series Mattresses 12 Part No 1148138Entrapment May OccurWARNINGPatient entrapment with bed side rails may cause injury or death
3 SAFETYPart No 1148138 13 Invacare®MA80 Series MattressesFire HazardElectricalDANGERSmokingDO NOT SMOKE while using this device. This system uses ro
4 OPERATIONInvacare®MA80 Series Mattresses 14 Part No 11481384 Operation4.1 Installing the Mattress Replacement SystemCAUTIONThe control unit and m
4 OPERATIONPart No 1148138 15 Invacare®MA80 Series Mattresses1. Remove the original foam mattress from the bed.2. If necessary, lower side rails to fa
4 OPERATIONInvacare®MA80 Series Mattresses 16 Part No 11481384.3 Installing the Control Unit1. Pull out the bed hook on the back of the control unit
4 OPERATIONPart No 1148138 17 Invacare®MA80 Series Mattresses5. Ensure the hose connector is flush with the control unit connector.FIGURE3 Connecti
4 OPERATIONInvacare®MA80 Series Mattresses 18 Part No 11481381. Examine the hospital grade power cord supplied with the control unit.2. Perform one of
4 OPERATIONPart No 1148138 19 Invacare®MA80 Series Mattresses4.6 Using the Front PanelFIGURE5 Using the Front PanelDETAIL “B” - MA85 FRONT PANELPow
Invacare®MA80 Series Mattresses 2 Part No 1148138© 2011 Invacare Corporation. All rights reserved. Republication, duplication or modification in whol
4 OPERATIONInvacare®MA80 Series Mattresses 20 Part No 1148138Power Button1. To turn the control unit on or off, press and release the Power button ( )
4 OPERATIONPart No 1148138 21 Invacare®MA80 Series MattressesAdjusting the Minimum Pressure Setting (A/P Low)The minimum pressure setting during Alter
4 OPERATIONInvacare®MA80 Series Mattresses 22 Part No 1148138Firm/Soft Buttons1. Select comfort pressure settings by pressing Firm ( ) or Soft ( ) but
4 OPERATIONPart No 1148138 23 Invacare®MA80 Series Mattresses3. Press and hold the Fowler ( ) and Max Inflate ( ) buttons.4. Raise the bed to 45°.5. P
4 OPERATIONInvacare®MA80 Series Mattresses 24 Part No 1148138Lock/Alarm Silence Button1. Press the Lock/Alarm Silence button ( ) to select one of the
4 OPERATIONPart No 1148138 25 Invacare®MA80 Series Mattresses4.8 Placing the Patient on the Mattress1. Press the Max Inflate button to turn on the con
4 OPERATIONInvacare®MA80 Series Mattresses 26 Part No 11481384.9 Transferring Patient From/To a Gurney1. Engage the wheel locks of the bed. Refer to t
4 OPERATIONPart No 1148138 27 Invacare®MA80 Series Mattresses4.11 Preparing for CPR Procedure1. Lift the connector cover (Detail “A”).2. Pull the hose
5 MAINTENANCE AND TROUBLESHOOTINGInvacare®MA80 Series Mattresses 28 Part No 11481385 Maintenance and Troubleshooting5.1 Cleaning the System1. Remove
5 MAINTENANCE AND TROUBLESHOOTINGPart No 1148138 29 Invacare®MA80 Series Mattresses8. Perform the following steps to clean the mattress cover:A. Using
CONTENTSPart No 11481383 Invacare®MA80 Series Mattresses1 GENERAL 4Symbols ...
5 MAINTENANCE AND TROUBLESHOOTINGInvacare®MA80 Series Mattresses 30 Part No 114813810. Perform the following steps to clean the mattress components (a
5 MAINTENANCE AND TROUBLESHOOTINGPart No 1148138 31 Invacare®MA80 Series Mattresses5.2 Changing the Filter1. Remove the two screws securing the filter
5 MAINTENANCE AND TROUBLESHOOTINGInvacare®MA80 Series Mattresses 32 Part No 11481385.3 TroubleshootingPROBLEM CAUSE SOLUTIONMattress not inflatingNot
NOTESPart No 1148138 33 Invacare®MA80 Series MattressesNOTES
NOTES Invacare®MA80 Series Mattresses 34 Part No 1148138NOTES
NOTESPart No 1148138 35 Invacare®MA80 Series MattressesNOTES
Invacare CorporationUSAOne Invacare WayElyria, Ohio USA44036-2125800-333-6900Canada570 Matheson Blvd E Unit 8Mississauga OntarioL4Z 4G4 Canada800-668-
1 GENERALInvacare®MA80 Series Mattresses 4 Part No 11481381 General1.1 SymbolsWarningsSignal words are used in this manual and apply to hazards or u
1 GENERALPart No 1148138 5 Invacare®MA80 Series Mattresses1.2 Limited WarrantyPLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL
2 OVERVIEWInvacare®MA80 Series Mattresses 6 Part No 11481382 Overview2.1 Label LocationDANGER-EXPLOSION HAZARD: DO NOT use in the presence of flamma
2 OVERVIEWPart No 1148138 7 Invacare®MA80 Series Mattresses2.2 Typical Product ParametersElectrical ParametersPerformance ParametersMA80/MA85INPUT VOL
2 OVERVIEWInvacare®MA80 Series Mattresses 8 Part No 1148138Mechanical ParametersPATIENT COMFORT CONTROL PRESSURES:SOFT PRESSURE:FIRM PRESSURE:8 ± 4 mm
2 OVERVIEWPart No 1148138 9 Invacare®MA80 Series MattressesEnvironmental ParametersMA80/MA85OPERATING CONDITIONSAMBIENT TEMPERATURE:RELATIVE HUMIDITY: