Owner’s Operation and Maintenance Manual
DEALER: This manual MUST be given to
the user of the tub.
USER:
BEFORE using this tub, read this
manual and save for future reference.
For more information regarding
Invacare products, parts, and services,
please visit www.invacare-ccg.com
Supine Bath Tub
IH6300 Fixed Height
IH6300ADH Adjustable Height
AFTER INSTALLATION, PLEASE CALL CUSTOMER SERVICE AT
1-800-668-2337 TO ARRANGE FOR AN IN-SERVICE ON THE TUB
Seitenansicht 0
1 2 3 4 5 6 ... 31 32

Inhaltsverzeichnis

Seite 1 - Supine Bath Tub

Owner’s Operation and Maintenance ManualDEALER: This manual MUST be given to the user of the tub.USER: BEFORE using this tub, read this manual and sav

Seite 2

SECTION 1—GENERAL GUIDELINESSupine Bath Tub 10 Part No 1096926Whenoperatingwaterpressureexceeds55p.s.i.,pressureregulators(similartothe¾‐i

Seite 3 - TABLE OF CONTENTS

SECTION 1—GENERAL GUIDELINESPart No 1096926 11 Supine Bath TubTheSupineTubisspecificallydesignedtobeusedwithabathlift.ICCGrecommendsuse

Seite 4

SECTION 2—DISINFECTING THE TUBSupine Bath Tub 12 Part No 1096926SECTION 2—DISINFECTING THE TUB WARNINGDisinfect the tub after EACH use to avoid resid

Seite 5 - SPECIAL NOTES

SECTION 3—OPERATIONPart No 1096926 13 Supine Bath TubSECTION 3—OPERATIONOverviewThetu bisdesignedtoimprovetheefficiencyandenvironmentalsafet

Seite 6 - LABEL LOCATION

SECTION 3—OPERATIONSupine Bath Tub 14 Part No 1096926Operating the Tub WARNINGDO NOT stand in tub. Tub surfaces are slippery when wet. Serious injury

Seite 7 - TYPICAL PRODUCT PARAMETERS

SECTION 3—OPERATIONPart No 1096926 15 Supine Bath TubOperating Shower/Massage WandNOTE:STEPS1‐3aretobeperformedwhileholdingtheshower/massa

Seite 8 - FEATURES

SECTION 3—OPERATIONSupine Bath Tub 16 Part No 1096926Adjusting Tub Height - Model IH6300ADH ONLY WARNINGDO NOT stand on tub frame. Frame is slippery

Seite 9 - SECTION 1—GENERAL GUIDELINES

SECTION 4—INSTALLATIONPart No 1096926 17 Supine Bath TubSECTION 4—INSTALLATIONPreparing the Work AreaNOTE:Forthisprocedure,refertoFIGURE 4.1.1.

Seite 10 - Operation

SECTION 4—INSTALLATIONSupine Bath Tub 18 Part No 1096926Securing the TubNOTE:Forthisprocedure,refertoFIGURE 4.1onpage 17andFIGURE 4.2.1. Set

Seite 11 - 11 Supine Bath Tub

SECTION 4—INSTALLATIONPart No 1096926 19 Supine Bath TubElectrical Hook-up WARNINGBe certain ALL electrical work is performed by a qualified electric

Seite 12 -  WARNING

Supine Bath Tub 2 Part No 1096926 WARNINGDO NOT USE THIS PRODUCT OR ANY AVAILABLE OPTIONAL EQUIPMENT WITHOUT FIRST COMPLETELY READING AND UNDERSTANDI

Seite 13 - SECTION 3—OPERATION

SECTION 4—INSTALLATIONSupine Bath Tub 20 Part No 1096926FIGURE 4.3 Electrical Hook-upElectrical Hookup115 V ElectricalIH6300 Fixed HeightUse 15 Amp

Seite 14 - Operating the Tub

SECTION 4—INSTALLATIONPart No 1096926 21 Supine Bath TubElectrical Installation DANGERThe National Electric Code requires the use of a ground fault c

Seite 15 - Operating Shower/Massage Wand

SECTION 4—INSTALLATIONSupine Bath Tub 22 Part No 1096926Plumbing Hook-Up WARNINGBe certain ALL plumbing work is performed by a qualified plumber and

Seite 16 - Lowering the Supine Tub

SECTION 4—INSTALLATIONPart No 1096926 23 Supine Bath TubPlumbing Installation WARNINGBe certain ALL plumbing work is performed by a qualified plumber

Seite 17 - SECTION 4—INSTALLATION

SECTION 4—INSTALLATIONSupine Bath Tub 24 Part No 1096926FIGURE 4.6 Plumbing Installation and SchematicDETAIL “A” - PLUMBING INSTALLATIONCold Water I

Seite 18 - Securing the Tub

SECTION 5—MAINTENANCEPart No 1096926 25 Supine Bath TubSECTION 5—MAINTENANCE WARNINGBefore performing any maintenance to the tub, make certain that t

Seite 19 - Electrical Hook-up

SECTION 5—MAINTENANCESupine Bath Tub 26 Part No 1096926Every Six Months❑ Verifythewaterflowtemperaturewithaseparatethermometer.❑ Visuallyinsp

Seite 20

SECTION 5—MAINTENANCEPart No 1096926 27 Supine Bath TubHeavy Duty Cleaning Procedure WARNINGDO NOT use this product or any available optional equipme

Seite 21 - Electrical Installation

SECTION 5—MAINTENANCESupine Bath Tub 28 Part No 1096926Calibrating the Disinfectant Flowmeter WARNINGDO NOT use this product or any available optiona

Seite 22 - Plumbing Hook-Up

SECTION 5—MAINTENANCEPart No 1096926 29 Supine Bath Tub15. Adjusttheinjectorbyperformingoneofthefollowing:• BottomofFloatingBallReadsGrea

Seite 23 - Plumbing Installation

TABLE OF CONTENTSPart No 1096926 3 Supine Bath TubTABLE OF CONTENTSSPECIAL NOTES ...

Seite 24

SECTION 5—MAINTENANCESupine Bath Tub 30 Part No 1096926Maintenance RecordDATE MAINTENANCE PERSONTYPE OF MAINTENANCE PERFORMEDNOTES

Seite 25 - SECTION 5—MAINTENANCE

SECTION 6—TROUBLESHOOTINGPart No 1096926 31 Supine Bath TubSECTION 6—TROUBLESHOOTINGNOTE:Ifproblemsarenotremediedbythesuggestedmeansabove,c

Seite 26 - Glide Pads

LIMITED WARRANTYPLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975.Thi

Seite 27 - Heavy Duty Cleaning Procedure

TABLE OF CONTENTSSupine Bath Tub 4 Part No 1096926TABLE OF CONTENTSSECTION 5—MAINTENANCE ... 2

Seite 28

SPECIAL NOTESPart No 1096926 5 Supine Bath TubSPECIAL NOTESSignalwordsareusedinthismanualandapplytohazardsorunsafepracticeswhichcouldr

Seite 29

LABEL LOCATIONSupine Bath Tub 6 Part No 1096926LABEL LOCATION! WARNINGELECTRICAL SHOCK HAZARD. All disassembly and maintenance of the Supine Tub MUST

Seite 30 - Maintenance Record

TYPICAL PRODUCT PARAMETERSPart No 1096926 7 Supine Bath TubTYPICAL PRODUCT PARAMETERSIH6300FIXED HEIGHTIH6300ADH ADJUSTABLE HEIGHTHeight Tub Rim 3

Seite 31 - SECTION 6—TROUBLESHOOTING

FEATURESSupine Bath Tub 8 Part No 1096926FEATURESCONTROL PANELTUB INTERIORModel IH6300ADH Adjustable Height TubTub Fill ON/OFFDisinfectant Flow MeterD

Seite 32 - LIMITED WARRANTY

SECTION 1—GENERAL GUIDELINESPart No 1096926 9 Supine Bath TubSECTION 1—GENERAL GUIDELINES WARNINGSECTION 1 - GENERAL GUIDELINES contains important in